Systems for the examination of components present in a vessel or evolved from a process
Measures the concentration of gases and vapours in real time
All Hiden residual gas analysers are application tested and calibrated to provide the highest quality performance, and they are backed by a 3 year warranty and lifetime service support.
Overview
The Hiden RGA Series includes three specification levels :
HALO RGA - for residual gas analysis
Suitable for vacuum fingerprint analysis, leak detection and trend analysis
3F Series RGA - triple filter mass spectrometers for analytical applications
Triple filter technology provides extended mass ranges, increased mass resolution, enhanced sensitivity and improved contamination resistance for more demanding analytical applications.
3F-PIC PIC - pulse ion counting detection for fast event studies
Ultra-fast pulse ion counting detection provides seven decade continuous dynamic range with optimum sensitivity at UHV/XHV
Features
- Mass range options: 50, 100, 200, 300, or 510 amu.
- Detector options give sensitivity for high vacuum to ultra high vacuum, UHV.
- Stability (less than ±0.5% height variation over 24 h)
- MASsoft control via RS232, USB or multiple systems via Ethernet LAN
- Fast access for Histogram, Trend Analysis and Analog peak displays
- Mixed mode scanning e.g. Trend, Histogram and Analog peaks in multiple-windows
- Simultaneous real time display of graphical and tabular trend analysis data
- Cursor for peak height identification under mouse control
- Real time background subtract, Automatic mass scale alignment
- Data export facility to ASCII format and to all Windows devices
- netMASsoft, an on board web server for RGA operation from your Web browser
Specifications
Mass range |
50,100, 200, 300 or 510 amu |
Minimum detectable concentration (HALO) |
2X10-13 mbar |
Minimum detectable concentration (3F) |
2X10-14 mbar |
Minimum detectable concentration (3F-PIC) |
2X10-15 mbar |
Ion Source |
Range of electron impact ion source types available |
Ion Source control |
Yes |
Electron emission |
Software variable 1µA to 2mA |
Electron energy |
Software variable 0V to 150 eV |
Ion energy |
Software variable 0V to 10 eV |
Detector |
Faraday cup and electron multiplier |
Mounting flange (HALO) |
DN-35-CF (2.75"/70mm OD Conflat type) |
Mounting flange (3F – Triple Filter) |
DN-63-CF (4.5”/114mm OD Conflat type) |