A specialist gas analysis system for fast event transient analysis
Measures the concentration of fast event gases and vapours in real time
The Hiden HPR-20 QIC TMS Transient MS is configured for fast event gas analysis of gases and vapours at pressures near atmosphere. Ideal for fast gas switching experiments the MS features the Hiden QIC quartz-lined 0.9 m sampling interface.
Overview
The inlet, operating at 200 °C, provides response times of less than 150 ms to changes in gas composition with a 5 decade response time in < 200 ms.
The QIC inlet is coupled directly to Hiden’s Pulse Ion Counting (PIC) digital MS which is capable of measurement speeds of up to 500 data points/s over the entire 7 decade dynamic range.
Features
- 0.9 m fast response QIC capillary - 150 ms response time
- Open ion source and optimised pumping configuration for fast response
- Digital PIC detector - 7 decades continuous log scale
- Detection of low ppm to high % levels in < 200 ms
Specifications
Mass range |
200 or 300 amu |
Minimum detectable concentration |
5 ppm |
Maximum detectable concentration |
100 % |
Sample inlet pressure (Standard) |
100 mbar to 2 bar abs |
Response time to gas concentration change |
200 milliseconds (5 decades) |
Fast measurement speed |
Up to 500 measurements per second (7 decades) |
Sample flow required |
Options for < 1 atm cc/min to 20 atm cc/min |
Corrosive gas sampling options |
Yes |