Tip Effects

If the tip is broken, or contaminated, you can often get a "double tip" effect, where both the tip and the contamination are scanning the surface. See the example below.

AFM artifact due to a double tip

In the example above, each DNA molecule has a "twin", caused by the double tip.

How to avoid it

You can check that this is occurring with a tip characterisation sample. If it happens, change the tip. Once a tip is broken, you must change it. If it is contamination, you MIGHT be able to clean it. But most often, even in this case, you will have to change to a new tip.

This artifact is common after scanning for a long time with the same tip. Tips may last very long, or sometimes go blunt quickly. Some samples (known as "tip-eaters") have a tendency to lead to blunt tips very quickly.
In order to find out if your tip is blunt, scan a well-known sample like this one.

AFM artifact due to a blunt tip

In the example above, the image on the right is made with a blunt tip, the left with a sharp (new) one.

How to avoid it

Use your tips carefully! If it gets blunt, replace it! In general, scanning with a lower applied force (see setpoint) helps to reduce tip wear.

Some low-wear tips which can scan for longer before becoming blunt are also available.

If the tip is dirty or contaminated (often with parts of the sample sticking to it), you can get strange repeating shapes in the image. Sometimes it is obvious that this has occurred, but sometimes it is not. In the example below there are two images of the same sample, one taken with a clean tip, and one with a badly contaminated tip.
Using a sample you already know can help you diagnose this problem. There are a number of tip-characterisation samples available.

AFM artifact due to a dirty or contaminaed tip

 

In the example above, the image on the right was scanned using a very badly contaminated tip, the one on the left with a clean tip, but the sample is the same!

This shows the drastic effects than can occur when the tip is dirty.


How to avoid it

Change the tip. Once a tip is broken, you must change it. If it is contamination, you might be able to clean it. But most often, even in this case, you will have to change to a new tip.

Note that the image above is of a sample of BOPP film (Biaxially Oriented PolyPropylene). The use of this sample in tip characterization, and even to clean the tip has been discussed by Nie et al in:
H.-Y. Nie, M. J. Walzak, and N. S. Mcintyre, "Use of biaxially oriented polypropylene film for evaluating and cleaning contaminated atomic force microscopy probe tips: an application to blind reconstruction," Review of Scientific Instruments, vol. 73, pp. 3831-3836, 2002.

However, you could use any well-known sample to characterise the state of your tip. The advantage of BOPP is that it is also possible to clean the tip by performing indentation into the sample. In principle you could use other samples for this. It's worth remembering, however that you are very unlikely to be able to return to the pristine state of a new tip, by any cleaning method.

 

This is an inherent feature of AFM and can never be fully removed. Any AFM image is a convolution of the shape of the probe, and the shape of the sample. This has the effect of making protruding features appear wide, and holes appear smaller (both narrower and often less deep, too). Broader (less sharp) probes will enhance the effect, as shown below.

 

AFM artifact due to tip-sample convolution

Illustration of convolution of AFM probe and sample giving rise to the image (red).

 

Probe-sample convolution tends to have the greatest effect on features of similar or smaller radii than the probe. it can be reduced by deconvolution techniques.