The cantilever
is the most common sensor of the force interaction in atomic force
microscopy. The atomic force microscope acquires any information about
a surface because of the cantilever beam mechanical deflections which
are detected by an optical system. In noncontact microscopy, resonators
of the tuning fork-type are frequently used. Such sensors require
tracking of the resonance frequency shift upon the probe-surface
interaction onset.
Normally, cantilever is a beam in the form of a rectangular parallepiped (Fig. 1a) having length , thickness ( ) and width ( ) or in the form of two beams connected at an angle (Fig. 1b) having a tip with length at its free end. Let us examine below the rectangular cantilever in detail. Its characteristic dimensions are shown in Fig. 1a. The probe's tip interacts with the surface. Assume that the point force acting from the sample is applied to the tip's apex.
 |
 |
| Fig. 1a. Rectangular cantilever with a tip. |
Fig. 1b. V-shaped cantilever with a tip. |
The force acting on
the probe has sometimes not only vertical but also horizontal
components. Therefore, the cantilever tip can deflect not only along
the -axis but in two other directions: and (see Fig. 1a). Let's call the force vertical component the normal force and longitudinal and transverse components – the lateral forces.
Because in AFM the
tip-sample interaction influences the cantilever deformation, to
determine the force one should know the cantilever deformation
stiffness in various directions. Consider that the tip deflection
vector (having components , , ) is linearly dependent on the applied force in accordance with the Hooke's law [1]:
 |
(1) |
The "constant" of proportionality is the second rank tensor which we call the inverse stiffness tensor. It contains all the information about elastic properties of the cantilever.
To find the components of the tensor
it is necessary to solve the problem of the cantilever static
deformation under the influence of forces directed along different
axes. For the sake of clarity, we write the formula (1) as a matrix
expression:
 |
(2) |
Notice that the
optical system detects not tip deflection but inclination of the
cantilever top surface near its free end. Two angles are measured:
deflection of the normal from vertical in the plane (angle ) and in the orthogonal direction – in the plane (angle ).
Instead of (2), we can write for the mathematical convenience the matrix expression relating angles and directly with force components.
 |
(3) |
The introduced matrix, however, does not contain full information about the cantilever elastic properties in contrast to tensor .
Summary.
- The information about a sample in AFM
can be obtained only from the cantilever deformation. The optical
system allows to measure two angles defining the cantilever top plane
inclination.
- To determine the force acting on
the cantilever one should know its elastic properties which are
described by the second rank tensor of the cantilever inverse stiffness.
- The deformation-force relation is modeled by the linear Hooke's law written as the tensor expression.
References
- Handbook of Micro/Nanotribology / Ed. by Bhushan Bharat. - 2d ed. - Boca Raton etc.: CRC press, 1999. – 859 c.
|