| I(z) Spectroscopy |
|
|
| STM | |
|
The I(z) Spectroscopy is related to LBH spectroscopy
and can be used for providing an information about the z-dependence of
the microscopic work function of the surface. Next important use of the
I(z) Spectroscopy is concerned with for testing of the STM tip quality.
The tunneling current IT in STM exponentially decays with the tip-sample separation z as IT ~ exp(-2kz), where the decay constant is given by 2k = 2(2mU/h2)1/2
U is the average work function Uav = (Us + Ut )/2, where Ut and Us are the tip and sample work functions, respectively. References
|
|