SFM can be used not only as tool for topography
acquisition but also can be used to produce spatially resolved maps of
the surface or material properties of a sample; these include charge
density, adhesion and stiffness,
as well as the force required to break specific igand–receptor bonds.
SFM can also be used as a tool for force spectroscopy – measuring
forces as a function of distance. For oscillating cantilever tip-sample
force can affect some other characteristics of cantilever oscillation –
amplitude, frequency, phase, dissipation
etc. Correspondingly dependence of these characteristics upon
tip-sample distance can also be regarded as spectroscopic data.
Force is measured in an SFM by collecting a force curve, which is a plot of cantilever deflection, dc,
as a function of sample position along the z-axis (i.e. towards or away
from the probe tip; the z-piezo position). It assumes a simple
relationship (i.e. Hooke’s Law) between the force, F, and the
cantilever deflection:
F = - k dc
where k is the spring constant of the cantilever.
Some other forces included in tip-sample interaction under dc approach
or retracting motion are presented on the figure left [1]. Used definitions see below.
The
interpretation of AFM force curves relies almost entirely on
established force laws, particularly those determined using the SFA [2].
These force laws describe force as a function of the probe–sample
separation distance (D) rather than as a function of the z-piezo
position. Thus, to be useful, the force curves must be transformed into
descriptions of force as a function of distance, F(D). However, current
SFMs do not have an independent measure of D. Instead, the
transformation to D is achieved by subtracting the cantilever
deflection from the z-piezo movement.
For a very hard surface,
zero separation is defined as the region in the force curve in which
the cantilever deflection is coupled 1:1 with the sample movement; this
appears in the force curve as a straight line of unit slope. A
corrected curve is called a force–distance curve. Notice that
determining D by this approach requires that the tip make contact with
the sample. In practice, there are two factors (long-range forces and
sample elasticity) that can make determining the point of contact very
difficult. A complete force curve includes the forces measured as the
probe approaches the sample and is retracted to its starting position.
Because the forces on the tip can vary as it is moved toward or away
from the sample, for the purposes of presentation, we will divide the
force curve into approach and retraction portions and consider them
separately.
Definitions
A Hamaker constant
a Monomer length
D Probe–sample separation distance
E Elastic modulus
k Boltzmann’s constant
L Brush thickness in a good solvent
L* Inverse Langevin function
N Number of units in polymer
R Radius of probe sphere
s Mean distance between polymers
T Absolute temperature
U Bond energy
x Elongation of polymer
d Indentation depth
e Dielectric of the medium
g Surface energy between tip and sample
gL Surface energy of the liquid
n Poisson ratio
L Characteristic length of bond
l Debye length of the medium
q Angle related to the geometry of
the tip–sample contact
sR Surface-charge density of sphere
sSSurface-charge density of sample
t Period over which the bond will
rupture
t0 Reciprocal of the natural bond
frequency
References
TIBTECH 17, 143 (1999).
Israelashvili, J.N. (1992) Intermolecular and Surface Forces, Academic Press.